COMPOSITION OF AN ETC EVO II :
A tank and a lid inert to acids, an integrated heating device, a plate for containers.
Decontamination and conditioning using acid vapour condensation:
This hot decontamination procedure, using an extremely pure acid vapour condensate, and without interaction with laboratory air, prevents contamination from entering into the containers, which are treated identically in this way. It is ideal for analysing trace and ultra-trace elements
For processing the following:
Materials: PFA, PTFE, PP, quartz, glass, etc.
Vessel: tubes for mineralization, autosampler cups, microwave reagents, storage bottles, pipette tips or cones, UV cuvettes, quartz/glass tubes for NMR.
Economical:
Quality reagents for analyses, purified by evaporation on each wash cycle, are used in small quantities (400-500 ml) several times. This therefore allows you to make significant savings in your laboratory budget, cost of purchasing or producing reagents (purification), cost of reprocessing acid effluents.
Temperature controller for ETC EVO II or CRD system sold separately (see sections “Controllers” and “CRD system” in the category “Clean and/or condition the vessel”).
Our devices have passed the acceptance tests in class 1 clean rooms and satisfy the standards EC EMC 89/336/EEC – LVD 73/23/EEC – EN 61010
Operating principle
The acid vapor, very pure and hot, emitted from the bath, enters the containers by convection. As it condenses on the inner wall of the containers, it dissolves any contaminants present on this wall and in the microfissures. The droplets containing these contaminants and impurities are returned to the tank. The contamination extracted in this way is then sequestered in the acid bath.
The operating temperature of the ETC is 110°C (maximum).
Vessel support
The vessel is placed in the orifices machined to the dimensions of the containers on the PTFE plate.
Economical
Quality reagents for analyses, purified by evaporation on each wash cycle, are used in small quantities (400-500 ml) several times. This therefore allows you to make significant savings in your laboratory budget (cost of purchasing or producing reagents, cost of reprocessing acid effluents).
Temperature controller or CRD system sold separately (see sections “Controllers” and “CRD system”).
Plate included in your ETC EVO II
OPTIONAL : Closure plate and P’Tips Rack